首頁>產(chǎn)品中心>日本光明北川式氣體檢測管>氣體快速檢測管>120UH硫化氫H2S氣體觀測管120UH HYDROGEN SULPHIDE 2-20%

硫化氫H2S氣體觀測管120UH HYDROGEN SULPHIDE 2-20%

簡要描述:

硫化氫H2S氣體觀測管120UH HYDROGEN SULPHIDE 2-20%,為您提供硫化氫檢測管的詳細(xì)介紹和價(jià)格,我公司為提高產(chǎn)品質(zhì)量投資上百萬元建立了檢測管自動(dòng)化生產(chǎn)線,引進(jìn)*的標(biāo)準(zhǔn)氣配置裝置,建立了嚴(yán)格的檢測管質(zhì)量保證體系,并取得了計(jì)量器具生產(chǎn)許可證。日本光明理化學(xué)工業(yè)株式會(huì)社在氣體分析領(lǐng)域開拓了六十年,形成了于世的“北川"式氣體快速分析法,“北川"的歷史就是檢測管的歷史。

更新時(shí)間:2026-03-11

發(fā)郵件給我們:554328895@qq.com

分享到: 1
在線留言
硫化氫H2S氣體觀測管120UH HYDROGEN SULPHIDE 2-20%

硫化氫H2S氣體觀測管120UH HYDROGEN SULPHIDE 2-20%

Tube 20UH
1. PERFORMANCE
1) Measuring range :2-20%
Number of pump strokes 1/2(50ml)
2) Sampling time :1 minute/1/ 2 pump stroke
3) Detectable limit :0.05%
4) Shelf life :3 years
5) Operating temperature :0~40℃
6) Reading :Direct reading from the scale calibrated by 1 /2 pump stroke
7) Colour change :Pale blue→Black
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:5% RSD-high:5%
3. CHEMICAL REACTION
By reacting with Cupric sulphate(II), Cupric sulphide is produced.
H2S+CuSO4→CuS
4. CALIBRATION OF THE TUBE
STANDARD GAS CYLINDER METHOD
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump
20
18
16
14
12
10
8
6
4
2
50ml
%
????
????
????
????
????
????
120UH HYDROGEN SULPHIDE
(NOTE)
1. Where the discolouration is over the scale, non-reacted Hydrogen sulphide remains in the pump. This residual
Hydrogen sulphide is not exhausted even if the pump handle is pushed back, and it still remains in the pump.
When the handle is pulled next, it is exhausted toward the operator from the stopper side of the pump.
Accordingly, air purging of the pump inside by pulling/pushing the handle several times should be done in
fresh air where there is not Hydrogen sulphide, because Hydrogen sulphide is very toxic. During the air
purging, protection apparatus such as a gas mask should be worn.
2. Maximum sampling volume of the gas shall be approximay 70mr.
Substance Interference Coexistence
Sulphur dioxide FIG.1 The accuracy of readings is not affected. Lower readings are given.
Carbon dioxide The accuracy of readings is
not affected.
0 2 4
20
Sulphur dioxide〔%〕?
FIG.1 Influence of Sulphur dioxide